Abstract

The electrothermal atomic absorption spectrometric determination of cadmium and lead in biological certified reference materials (CRMs) has been carried out by using NH 4H 2PO 4, Ni, Pd, Ni+NH 4H 2PO 4, Pd+NH 4H 2PO 4 and Ni+Pd+NH 4H 2PO 4 as chemical modifiers. A comprehensive comparison was made among the modifiers in 1% Triton X-100 plus 0.2% nitric acid as diluent and without modifier. Zeeman background correction and graphite tubes inserted with platforms were used. Comparison was made in terms of pyrolysis and atomization temperatures, atomization and background absorption profiles. Ni+Pd+NH 4H 2PO 4 modifier mixture was found to be preferable for the determination of Cd and Pb. Pyrolysis temperatures of analytes were increased up to 900 °C for Cd and 1250 °C for Pb by using Ni+Pd+NH 4H 2PO 4 in 1% Triton X-100 plus 0.2% nitric acid diluent solution. Biological CRMs were analyzed to verify the accuracy and precision of this method. Depending on the biological sample type, the percent recoveries were increased from 62 to 102% for Cd and from 58 to 106% for Pb by using the proposed modifier mixture. The detection limits of Cd and Pb were found to be 0.04, 0.92 μg l −1, respectively.