Abstract

A simple continuous flow generation of volatile bromine is described for the determination of low concentrations of the elements by atmospheric-pressure argon microwave induced plasma (MIP) surfatron. Bromine is continuously generated by mixing the bromide with sulphuric acid and hypochlorite solutions. The bromine vapor is separated from the aqueous phase by a gas—liquid separator and is desiccated by passing it through concentrated sulphuric acid. The detection limit attained was 2 μg/l. and the precision was ±0.7% (at the 80 μg/l. level). The proposed determination is very selective if oxidizing/reducing agents are absent. The procedure has been tested for bromide determination in two drug preparations. Good agreement between the experimental results and the certified values has been obtained.

Full Text