Abstract

The bismuth contents of various digested urine samples and prescription medicines were determined by atomic absorption spectrometry combined with hydride generation. The procedure followed was a standard addition method for urine and direct calibration for the prescription medicines. The detection limit of the method was determined to be 320 pg ml-1 Bi with an analytical frequency of 150 h-1. A relative standard deviation of 4.7% was found for Bi in urine at the level of 4.3 ng ml-1 Bi. Interference caused by NiII, CoII, CuII, AgI, SeIV, SbIII and HgII could be controlled with a masking solution of thiourea (0.2%)-KI (10%).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.