Abstract

Mixtures of bioactive peptides have been analyzed by capillary zone electrophoresis/mass spectrometry (CZE/MS) using an on-line coaxial continuous-flow fast atom bombardment interface. High separation efficiencies (up to 410,000 theoretical plates) were obtained from low femtomole levels of peptides. The analysis of basic peptides was accomplished by using aminopropyl-silylated CZE columns to minimize zone broadening due to adsorption effects. CZE/MS/MS data were acquired from femtomole levels of peptides in electrophoretic real time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.