Abstract
In this paper, vortex-assisted extraction using the ultraperformance liquid chromatography analysis method was performed to determine benzo[a]pyrene in camellia oil. Optimum results were obtained when 0.5 g of oil sample was used followed by vortex-assisted extraction for 10 min with 25 mL of acetonitrile. Chromatographic separation was performed on an Agilent ZORBAX Eclipse Plus C18 column (2.1mm×100mm, particle size 1.8 μm). The optimum mobile phase comprised 70% acetone and 30% water. The detection limit of benzo[a]pyrene was 0.2 μg/kg. The recoveries were in the range of 81.0-97.0%. The proposed method was simple and fast, and it provided high throughput in the determination of benzo[a]pyrene in an oil matrix sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.