Abstract

Abstract A high-performance liquid chromatographic (HPLC) method for the determination of individual B6 vitamers was developed, and its application to dog, rat, monkey and human serum is described. Serum samples were deproteinated with 8% aqueous perchloric acid and the extracts were purified using an anion-exchange column. Analytical separation of the B6 vitamers and the internal standard (4′-deoxypyridoxine) was achieved by isocratic elution with a nitrile phase column in less than 30 minutes. The described methodology is simple and fast, and yields high percentage recoveries (> 90) of the vitamers with low inter- and intra-day variations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.