Abstract

An analytical method is described for the determination of the avermectins in plasma based upon high-performance liquid chromatography of flourescent derivatives of these compounds. The analyte is isolated by adsorption chromatography on Florisil, dehydrated in an acetic anhydride-pyridine mixture, and the fluorophore is further separated by chromatorgraphy on silica gel in advance of introduction into a reversed-phase system. This method, which can be applied to samples containing as little as 0.2 ng drug per ml, has an accuracy of 5% mean relative error and a precision of 8% relative standard deviation. A study and discussion of several factors which affect the analytical reaction are included.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.