Abstract

Arsenic in homeopathic drugs was determined by coupling a volatile generation with inductively coupled plasma-atomic emission spectrometry. The method is based on the chemical vaporization of arsenic(III) with bromide ions in sulfuric acid media using a batch procedure and subsequent introduction of the gaseous analyte into the plasma torch. The main and interactive effects of the experimental variables affecting this method were evaluated by a 2-level full factorial design. In optimized conditions by Simplex, the method shows an absolute detection limit (3 s) of 0.28 microg for the injection of 230 microL sample. The precision (% relative standard deviation) of the determination was 4.2% at a level of 50 microg/mL As(III) (n = 5). The interference effect of various ions on the arsenic signal was evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.