Abstract

A graphite-furnace AAS method using the stabilized-temperature platform furnace (STPF) concept, mixed palladium and magnesium nitrates as chemical modifier and Zeeman background correction has been applied to the direct determination of As, Cd, Pb and Se in highly mineralized waters used for medicinal purposes. These contain 20–40 g/l. concentrations of salts, mainly sodium and magnesium chlorides, bicarbonates and sulphates. The use of a pre-atomization cool-down step to 20° in the graphite-furnace programme reduced the background absorption. Increasing the mass of magnesium nitrate modifier to 5 times that originally proposed improved the analyte peak shape. Under these conditions, no interference was found in analysis of the chloride/bicarbonate type of water, but the sodium and magnesium sulphate type of water had to be diluted, and even then an interference remained. Calibration with matrix-free standard solutions was used, but use of spike recovery is strongly recommended for testing the accuracy. The limits of determination (4.65σ) of the proposed method for undiluted samples are 2.0 μg/l. for As, 0.05 μg/l. for Cd, 1.0 μg/l. for Pb and 1.5 μg/l. for Se.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.