Abstract

A procedure for the separation and determination of the five arsenic species As(III), As(V), monomethylarsonic acid, dimethylarsinic acid and arsenobetain by ion chromatography coupled to inductively coupled plasma mass spectrometry under the use of anion exchange is described. It has been successfully applied to the analysis of groundwater and extracts of contaminated soils. No interference of 40 Ar, 35 Cl and 75 As was observed when natural water samples were analyzed. The procedure shows good repeatability and accuracy in addition to easy handling. Detection limits for the arsenic species are in the range 0.4–0.8 μg l −1. A detection limit of 4 μg l −1 was obtained for the selenium species. Interference by Cl − up to concentrations of 400 mg l −1 can be neglected. The influence of several parameters (eluent concentration, amount of methanol in the mobile phase, flow rate) has been studied so as to optimize the best chromatographic conditions. The procedure has prospects for environmental biomonitoring using a sensitive and efficient determination of arsenic and selenium species, as shown for groundwater samples spiked at the 50 μg l −1 level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.