Abstract

The aim of this study was to develop a simple and fast ultrasound-assisted ionic liquid dispersive liquid-liquid phase microextraction (UA-IL-DLLME) method for preconcetration of trace antimony and tin in beverage samples. The novelty of this study was based on the application of ligandless UA-IL-DLLME using low-density ionic liquid and organic solvents for preconcentration of Sb and Sn. The concentration of Sb and Sn were quantified using ICP-OES. Under the optimum conditions, the calibration graph was found to be LOQ-250µgL−1 (r2=0.9987) for Sb and LOQ-350µgL−1 for Sn. The LOD and LOQ of Sb and Sn ranged from 1.2to 2.5ngL−1 and 4.0 to 8.3ngL−1, respectively, with high preconcentration factors. The precisions (%RSD) of the proposed method ranged from 2.1% to 2.5% and 3.9% to 4.7% for Sb and Sn, respectively. The proposed method was successfully applied for determination of Sb and Sn in beverages.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.