Abstract
The role that an ultrasonic piezoelectric transducer plays in an ultrasonic measurement system can be described in terms of the transducer's input electrical impedance and its sensitivity. Here, a new model-based approach is proposed to determine both the transducer impedance and sensitivity in a pulse-echo setup. This new method is much simpler to apply than previous "self-reciprocity" calibration methods for determining sensitivity and generalizes those methods. It is demonstrated that sensitivities obtained with this new method agree well with the sensitivities obtained by a three-transducer method commonly used in calibration studies. It is demonstrated that at the megahertz frequencies at which ultrasonic transducers operate it is important to compensate for cabling effects in these measurements. The influence of the pulser/receiver settings on the results obtained also will be discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.