Abstract

A fast, sensitive, simple, and highly reproducible method for routine assay of ammonium ion (NH4+) was developed by using HPLC equipment. The method is based on the reaction of NH4+ with o-phthalaldehyde (OPA) in the presence of 2-mercaptoethanol. After an on-line derivatization, the resulting NH4(+)-OPA product was quantified by using fluorometric or spectrophotometric detection. For fluorometric detection, the excitation and emission wavelengths were 410 and 470 nm, respectively. The spectrophotometric detection was made by measuring absorbance at 410 nm. Results on the effects of OPA-reagent composition and pH, reaction temperature, sample matrix, and linearity of the assay are presented. Even though it took about 2 min from the time of sample injection to the appearance of sample peak, sample injections could be overlapped at an interval of about 1 min. Thus, the actual time needed for analysis was about 1 min per assay. The method can be used in a fully automated mode by using an autosampler injector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.