Al x Ga 1− x As Auger sensitivity factors have been determined by using PHI610 scanning Auger microprobe with pure elemental standards Al, Ag and matrix GaAs. The quantitative results of Al x Ga 1− x As measured by the present method are in very good agreement with X-ray double crystal measurements. It is shown that by using sensitivity factors obtained from the self-instrument, the accuracy of the quantitative AES analysis can be considerably improved compared with that using elemental relative sensitivity factors given by the PHI handbook or internal standard method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.