Abstract
This study established a method to determine the acrylonitrile in solid waste by automatic headspace gas chromatography. This method adopted the automatic headspace sampling, separated by DM - FFAP chromatographic column, detected by FID detector. The conditions of headspace balance time and equilibrium temperature were optimized. The linear range, detection limit, recovery, sensitivity and stability were tested. Under the conditions of 80°C and balance time of 30min, the method detection limit of acrylonitrile was 1.66μg /kg. The relative standard deviation was under 5%. The standard addition recovery was 92.7% ∼ 101.5%. The results showed that the before processing of this method is simple, rapid, no consumption of organic solvents, accurate and applicable. The method can meet the requirements of Identification standards for hazardous wastes-Identification for toxic substance content.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.