Abstract

This paper reported a novel headspace gas chromatographic (GC) technique on quantification of acidic and basic species. It is based on an acid–base reaction between the measured species and bicarbonate in an aqueous solution, which generates carbon dioxide in a closed headspace sample vial. By operating at 60 °C, carbon dioxide is completely released to the headspace and thus can be measured by GC with a thermal conductivity detector. Bicarbonate concentrations of 0.030 and 0.0025 mol/L are recommended for general applications and very small species content, respectively. This method is able to accurately measure small sample sizes (down to few milligrams or microliters). The present method is simple, accurate, and automatic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.