Abstract

A comparison of four of the most commonly used dosimetry protocols for the determination of absorbed dose to water in therapeutic kilovoltage x-rays using an ionization chamber (IAEA TRS-277, IPEMB, DIN and NCS) has been carried out. Owing to the different energy ranges and HVLs recommended by each protocol, backscatter factors, water-to-air mass energy absorption coefficient ratios and perturbation correction factors have been recast to a common quality range that all protocols satisfy individually to make a comparison possible. The results of the comparison show that in the sometimes reduced quality range originally included by the different protocols, determinations of absorbed dose to water at all beam qualities agree to within ±1.0% with that obtained using the second edition of the IAEA TRS-277 code of practice (1997). The extrapolation of data to a common beam quality range practically preserves the agreement for all the protocols except for that issued by the NCS at the extremes of the range, where differences of up to 1.8% and 1.4% have been found for low and medium energies respectively. In all cases the DIN protocol yields very good agreement with TRS-277.