Abstract
To establish a method for the measurement of 5-fluorouracil (5-FU), and tegafur (FT) in tear samples from patients treated with oral fluoropyrimidine anticancer agent S-1. Cross-sectional study. High performance liquid chromatographic (HPLC) method reported for plasma samples was modified for tear samples. Simulated-tear solutions containing lactoferrin, lysozyme and standard solution containing 5-FU or FT were prepared for preliminary measurements. Tear samples from seven patients treated with S-1 were included. The tears were collected following S-1 administration using Schirmer's strips. 5-FU and FT concentrations of pure standard solution without extraction process were detected as original concentration. However, on extraction, FT samples in simulated-tear solution showed a peak for 5-FU but not for FT. FT was converted to 5-FU in the extraction process. Decomposition from FT to 5-FU was suppressed when 50mg/mL bovine serum albumin was added during extraction. The mean concentrations of 5-FU and FT in tears during S-1 treatment were 0.17 ± 0.11 and 1.94 ± 0.71μg/mL, respectively. A simple HPLC method to determine 5-FU and FT in tear samples was established.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.