Abstract

The majority of long-lived radionuclides produced in the nuclear fuel cycle can be regarded as “difficult-to-measure” nuclides, hence chemical separation is needed before the nuclear measurement of them. A combined radiochemical procedure that enables the simultaneous determination of some “difficult-to-measure” nuclides in medium and low level radioactive wastes has been developed in our laboratory. Recently, this method has been extended for determination of 237Np and 93Zr. 237Np and 93Zr are pre-concentrated by co-precipitation on iron(II) hydroxide and zirconium oxide, separated by extraction chromatography using UTEVA, and measured by inductively coupled plasma mass spectrometry (ICP-MS). As even traces of polyatomic ions and isotopes at m/z 237 or 93 cause considerable interferences during ICP-MS detection, a purification step by extraction chromatography was needed. Analyzing real samples (evaporation concentrates of a nuclear power plant) 66–99% and 31–99% chemical yields were achieved for Np and Zr, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.