Abstract

Summary This paper presents a simple and accurate method for determination of uranium isotopes (234U and 238U) in seawater samples using alpha spectrometry. Uranium is pre-concentrated from seawater samples via adsorption on hydrotalcite at pH 6.5. The absorbent is dissolved into 50 ml of 8M HNO3; then filtered through a Dowex-1 column. U(VI) in the elution solution is reduced to U(IV) using zinc metal in 4M solution of HCl and then co-precipitated with LaF3. The chemical separation efficiency is found to be 97.12±2.68%, eliminating the need of using 232U tracer in other published methods. This method is validated via comparison with results obtained using inductively coupled plasma mass spectrometry (ICP-MS) and neutron activation analysis (NAA) for three seawater samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.