Abstract

Measuring 226Ra in urine at low levels is critical for both biomonitoring and radiological emergency response. Here we report a new analytical method to quantify 226Ra, as developed and validated by a simple dilute-and-shoot procedure, followed by Inductively Coupled Plasma-triple quadrupole-mass spectrometry detection using 'No Gas MS-MS' mode. The method provides rapid and accurate results for 226Ra with a limit of detection (LOD) down to 0.007ng/l (0.26Bq/l). This LOD is well below the recommended action levels for 226Ra detection in children and pregnant women (C/P) set by the Clinical Decision Guide (NCRP Report #161). Results for 226Ra obtained by this method are within ±7.0% of the target values of standard reference materials spiked in the urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.