Abstract
A high performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) based method has been developed and validated to confirm the presence of the synthetic growth promoter 17α-methyltestosterone (MT) in bovine serum samples. After a single-step extraction with acetonitrile, the enhanced determination of methyltestosterone was achieved by detecting its oxime derivative. In accordance with the criteria defined in Commission Decision 2002/657/EC and using deuterated testosterone (T-d3) as an internal standard, the method has been validated in terms of decision limit (0.05 ng/mL), detection capability (0.09 ng/mL), repeatability CV (7.04%), within-laboratory reproducibility CV (11.00%), and accuracy (87.00%).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.