Abstract

Activated carbon fiber (ACF) as extraction fiber for solid-phase microextraction (SPME) and its application for the analysis of banned organochlorine pesticides (OCPs) were investigated. Firstly, ACF was activated by different concentration of zinc chloride, which indicated that ACF activated by 60% zinc chloride had a reasonable specific surface area, pore volume and pore distribution. Secondly, the parameters for the ACF-SPME procedure, the adsorption and desorption conditions, were also optimized when coupled with gas chromatography-mass spectrometry (GC-MS). Thirdly, the ACF-SPME was used to analyze 17 kinds of OCPs in water. The linearity of most pesticides was found to be between 0.2 and 50 microg/l with GC-MS under the selected ion monitoring (SIM) acquisition mode. The limits of detection (LOD) at the sub microg/l were obtained. The work demonstrated here shows that ACF is a promising alternative for the SPME procedure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.