Abstract

Objective: To establish a automatic headspace gas chromatography-mass spectrometry (HS/GC-MS) method for the determination of 14 volatile organic compounds in urine. Methods: In September 2022, 10 ml urine sample was taken into a 20 ml headspace bottle, balanced for 30 min at 65 ℃, and then detected by HS/GC-MS and quantified by external standard method. Results: The 14 volatile organic compounds showed good linearity at 0.2-8.0 μg/L and 0.1-4.0 μg/L, with correlation coefficients ranging from 0.9956-0.9999. The recoveries were 79.8%-113.1% with relative standard deviations 0.05%-0.27% when three different concentration levels were added. Detection limit was 0.03-0.05 μg/L. Conclusion: The method is simple and convenient, and the recovery and precision meet the requirements. It can be used for the determination of common volatile organic compounds in urine.