Abstract
We have developed a new extraction method for the measurement of 129I by accelerator mass spectrometry (AMS) utilizing an anion exchange resin disk. In comparison to traditional methods such as solvent extraction and ion exchange, this method provides for simple and quick sample handling. This extraction method was tested on soil, seaweed and milk samples, but because of disk clogging, the milk samples and some of the seaweed could not be applied successfully. Using this new extraction method to prepare samples for AMS analysis produced isotope ratios of iodine in good agreement with neutron activation analysis (NAA). The disk extraction method which take half an hour is faster than previous techniques, such as solvent extraction or ion exchange which take a few hours. The combination of the disk method and the AMS measurement is a powerful tool for the determination of 129I. Furthermore, these data will be available for the environmental monitoring before and during the operation of a new nuclear fuel reprocessing plant in Japan.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.