Abstract
To develop a method for the determination of 11 perfluorinated compounds(PFCs) in drinking water by solid phase extraction-ultra performance liquid chromatography tandem mass spectrometry(UPLC-MS/MS). The target analytes in water samples were concentrated and purified by Oasis WAX SPE column and eluted with 5 mL methanol and 7 mL 0. 1% ammonia-methanol solution. The eluent was blown to near dryness under a nitrogen gas, diluting with methanol water solution(3∶7, V/V) to 1 mL for analysis by UPLC-MS/MS. There were good liner relationship in the range of 0. 5-50 μg/L. The detection limit and the limit of quantitation were 0. 03-0. 7 ng/L and 0. 5-3. 0 ng/L, respectively. The recoveries of 11 PFCs at three different spiked concentrations were 90. 0%-122. 3%, 87. 1%-130. 0% and 80. 0%-113. 7%, respectively, and the relative standard deviation were 1. 95%-8. 59%, 1. 27%-9. 08% and 2. 17%-10. 51%(n=6), respectively. This method has wide linear range, low detection limit and limit of quantitation, applicable for simultaneous analysis and monitorting of 11 PFCs in drinking water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.