Abstract

Objective: To establish a method for the determination of 1-methoxy-2-propanol in urine using headspace solid phase micro-extraction coupled with gas chromatography. Methods: The 1-methoxy-2-propanol was enriched by headspace solid phase micro-extraction fiber coated with carbene/polydimethylsiloxane (CAR/PDMS) . Single factor rotation method was used to optimize the conditions of extraction temperature, salt amount, and extraction time. The separation was performed on DB-5 (30 m×0.32 mm×0.25 μm) capillary column and detected with flame ionization detector. The quantification was based on the standard curve. Results: The concentration of 1-methoxy-2-propanol in urine was linear in the range of 0.50-10.0 mg/L, and the linear correlation coefficient was 0.9993. The detection limit of the method was 0.14 mg/L, and the limit of quantification was 0.45 mg/L. The recovery was 85.8% to 104.7%, and the RSD of intra- and inter-batch precision were 3.25%-6.65% and 0.81%-3.96%, respectively. Conclusion: The method is high sensitivity and simple operation, and is suitable for the determination of 1-methoxy-2-propanol in urine of occupational exposure population.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.