Abstract

A high-throughput mass spectrometric immunoassay (MSIA) system for the analysis of proteins directly from biological fluids is reported. A 96-well-format robotic workstation equipped with antibody-derivatized affinity pipet tips was used for the parallel extraction of specific proteins from samples and subsequent deposition onto 96-well arrayed matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOFMS) targets. Interferences from nonspecifically bound proteins were minimized through choice of appropriate affinity pipet tip derivatization chemistries. Sample preparation for MALDI-TOFMS was enhanced through the use of hydrophobic/hydrophilic contrasting targets, which also presented functionalities found to promote matrix/analyte crystal growth. Automated mass spectrometry was used in the unattended acquisition of data, resulting in an analysis rate of approximately 100 samples/h (biological fluid-->data). The quantitative MSIA of beta2m levels present in human plasma samples is given as illustration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.