Abstract

A simple and sensitive LC-MS method for the determination of periplocin in rat plasma was developed and validated. The chromatographic separation was carried out using a reverse-phase Kromasil C(18) column(150 × 4.6 mm, i.d., 5 µm) with a mobile phase composed of methanol-water (76:24, v/v). The flow rate of mobile phase was 0.8 mL/min. The calibration curve was linear within the concentration range 1-1000 ng/mL. The intra- and inter-day precisions across three validation days over the entire concentration range was lower than 9.2% in terms of relative standard deviation. Accuracy determined at three quality control concentrations ranged from -2.0 to 6.0% in terms of relative error. The validated method was applied to the pharmacokinetic study of periplocin in rat plasma after intravenous and intramuscular administration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.