Abstract
We developed and implemented an algorithm to determine and correct systematic distortions in low-energy electron diffraction (LEED) images. The procedure is in principle independent of the design of the apparatus (spherical or planar phosphorescent screen vs. channeltron detector) and is therefore applicable to all device variants, known as conventional LEED, micro-channel plate LEED, and spot profile analysis LEED. The essential prerequisite is a calibration image of a sample with a well-known structure and a suitably high number of diffraction spots, e.g., a Si(111)-7×7 reconstructed surface. The algorithm provides a formalism which can be used to rectify all further measurements generated with the same device. In detail, one needs to distinguish between radial and asymmetric distortion. Additionally, it is necessary to know the primary energy of the electrons precisely to derive accurate lattice constants. Often, there will be a deviation between the true kinetic energy and the value set in the LEED control. Here, we introduce a method to determine this energy error more accurately than in previous studies. Following the correction of the systematic errors, a relative accuracy of better than 1% can be achieved for the determination of the lattice parameters of unknown samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.