Abstract

van Weissenbruch R, Albers FWJ, Bouckaert S, Nelis HJ, Criel G, Remon JP, Suiter AM. Deterioration of the Provox™ silicone tracheoesophageal voice prosthesis: microbial aspects and structural changes. Acta Otolaryngol (Stockh) 1997; 117: 452-458.Device life of tracheoesophageal voice prostheses is limited due to deterioration of the polymers. A group of 55 postlaryngectomy patients fitted with a Provox™ voice prosthesis have been studied prospectively during 6 months. Thirty-seven prostheses were replaced due to a dysfunctional valve mechanism. Although colonization with Candida species was highly associated with destruction of the silicone material, other upper respiratory tract commensals, e.g. Staphylococcus aureus, were also demonstrated. Electron microscopy of the contaminated devices showed colonization and disruption of the silicone material by penetrating yeast hyphae. During the study a remarkable increase of intratracheal phonatory pressures was assessed with progressive colonization of the prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.