Abstract
Frequency-domain fluorometry relies on the measurement of the phase and amplitudes of the Fourier components of the time-dependent fluorescence signal. Experimental results that show that a conventional photomultiplier is subject to intensity-dependent phase shifts are presented. The measurements indicate that this is a problem well below the maximum linear current of the photomultiplier response. These results have important implications in frequency-domain fluorescence anisotropy experiments, in which the parallel and the perpendicular components of the emission intensity are inherently different from one another: a phase shift can be introduced by the photomultiplier.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.