Abstract

Organ dedicated PET devices provide improved imaging performance when compared to whole body systems. The present study summarizes the test carried out to study a new detector block designed for an organ dedicated PET system. This block includes three novel components namely the scintillator geometry and a retroreflector layer coupled to the entrance face, the photosensor and the readout electronics. We used arrays of 12 × 12 SiPM photosensors with 3 × 3 mm2 active area each and a pitch of 4.2 mm. We are proposing a new readout electronics that permits to reduce the 12 row and columns signals to only 8 without significant detector performance degradation. This approach also allows for resolving radioactive sources in the whole volume of the proposed crystal, significantly reducing the edge effect that typically rejects these events.An overall spatial resolution of about 1.8 mm FWHM is obtained for the whole scintillation volume, with an average energy resolution of 13% FWHM and a photon depth of interaction resolution (FWHM) of 3.7 mm.