Abstract
The BDPN-1K detector assembly for neutron radiation comprises the detector and a preamplifier connected by a 10-m-long cable. The detector can be disassembled. Before the sensitive surface of the diamond there is placed a cover with a converter disk cemented to it; this disk is a U/sub 3/O/sub 8/ layer applied to an aluminum foil. Since the detector can be disassembled, the disk can be replaced and other materials can be used as the converter. In order to determine the optimal thickness of the converter, the sensitivity of recording thermal neutrons with a particular detector assembly was measured and converters with various surface densities of /sup 235/U were used in the detector. Measurements were made in a standard field of thermal neutrons from a plutonium-beryllium neutron source inserted in a paraffin moderator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.