Abstract

Purpose: Developing photon-counting CT detectors requires understanding the impact of parameters, such as converter material, thickness, and pixel size. We apply a linear-systems framework, incorporating spatial and energy resolution, to study realistic silicon (Si) and cadmium telluride (CdTe) detectors at a low count rate. Approach: We compared CdTe detector designs with and pixels and Si detector designs with pixels of 30 and 60mm active thickness, with and without tungsten scatter blockers. Monte-Carlo simulations of photon transport were used together with Gaussian charge sharing models fitted to published data. Results: For detection in a 300-mm-thick object at 120kVp, the 0.5- and 0.225-mm pixel CdTe systems have 28% to 41% and 5% to 29% higher detective quantum efficiency (DQE), respectively, than the 60-mm Si system with tungsten, whereas the corresponding numbers for two-material decomposition are 2% lower to 11% higher DQE and 31% to 54% lower DQE compared to Si. We also show that combining these detectors with dual-spectrum acquisition is beneficial. Conclusions: In the low-count-rate regime, CdTe detector systems outperform the Si systems for detection tasks, whereas silicon outperforms one or both of the CdTe systems for material decomposition.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call