Abstract

A selective extraction of the correlated events due to three kinds of short-lived α-decay events,216Po (145 ms),217At (32.3 ms) and215Po (1.78 ms), from other random background events were examined by using a time interval analysis (TIA) from the viewpoint of detection sensitivity. In the theoretical treatment based on the probability and statistics, it was clarified that the detection sensitivity proportionally increased with shortening half-lives in the following order,215Po>217At>216Po. In the experiments, pulses from a liquid scintillator owing to α-decay events were analyzed using a multiple TIA technique and resulted in the highest sensitivity of219Rn−215Po pairs in the actinium series. Thus, the multiple TIA method was applied to the practical determination of219Rn−215Po pairs in natural samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.