Abstract

The ability of the Vitek AMS system to detect vancomycin resistance in enterococci was evaluated by comparing the results to an agar dilution screen method. Of 100 strains tested, 43 were resistant and 51 were susceptible by both tests. Two strains were intermediate by Vitek but susceptible by agar screen, one was intermediate by Vitek but resistant by agar screen, and four were susceptible by Vitek but resistant by agar screen. When the Vitek intermediate and resistant results were combined, the false resistant rate was 4.4% and false susceptible rate was 8.5%. In its current format, the Vitek system appears not to have acceptable accuracy for the detection of vancomycin resistance in enterococci.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.