Abstract

A hand-held device for laser-induced breakdown spectroscopy has been investigated for the determination of uranyl fluoride surface contamination. This research demonstrates the ability to successfully detect uranium on surfaces when using a low resolving power (λ/Δλ=4000) spectrograph, with a 5 mJ energy per 1 ns pulsed laser radiation, available as a commercially packaged hand-held system. Sand/uranyl fluoride mixtures are prepared to simulate residue likely encountered during decontamination efforts at facilities that handle uranium hexafluoride. Detection limits are described for four uranium lines with one revealing the capability to detect uranium at a level of 250 parts per million. Advantages of the studied compact device include that location specific information can be obtained on-site to augment contamination identification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.