Abstract

Gastric motility is coordinated by bio-electrical events known as slow waves. Abnormalities in slow waves are linked to major functional and motility disorders. In recent years, the use of high-resolution (HR) recordings have provided a unique view of spatiotemporal activation profiles of normal and dysrhythmic slow wave activity. To date, in vivo studies of gastric slow wave activity have primarily focused on the activation phase of the slow wave event. In this study, the recovery phase of slow waves was investigated through the use of HR recording techniques. The recovery phase of the slow wave event was detected through the use of the signal derivative, computed via a wavelet transform. The activation to recovery interval (ARi) metric was computed as a difference between the recovery time and activation time. The detection method was validated with synthetic slow wave signals of varying morphologies with the addition of synthetic ventilator and high frequency noise. The methods was then applied to HR experimental porcine gastric slow wave recordings. Ventilator noise more than 10% of the slow wave amplitude affected the estimation of the ARi metric. Signal to noise ratio below 3 dB affected the ARi metric, but with minor deviation in accuracy. Experimental ARi values ranged from 3.7-4.7 s from three data sets, with significant differences across them.