Abstract

To investigate whether sliding window adaptive histogram equalization (SWAHE) of digital mammograms improves the detection of simulated calcifications, as compared to images normalized by global histogram equalization (GHE). Direct digital mammograms were obtained from mammary tissue phantoms superimposed with different frames. Each frame was divided into forty squares by a wire mesh, and contained granular calcifications randomly positioned in about 50% of the squares. Three radiologists read the mammograms on a display monitor. They classified their confidence in the presence of microcalcifications in each square on a scale of 1 to 5. Images processed with GHE were first read and used as a reference. In a later session, the same images processed with SWAHE were read. The results were compared using ROC methodology. When the total areas AZ were compared, the results were completely equivocal. When comparing the high-specificity partial ROC area AZ,0.2 below false-positive fraction (FPF) 0.20, two of the three observers performed best with the images processed with SWAHE. The difference was not statistically significant. When the reader's confidence threshold in malignancy is set at a high level, increasing the contrast of mammograms with SWAHE may enhance the visibility of microcalcifications without adversely affecting the false-positive rate. When the reader's confidence threshold is set at a low level, the effect of SWAHE is an increase of false positives. Further investigation is needed to confirm the validity of the conclusions.