Abstract

The scope of this paper was to determine whether ischemic and reperfusion damage in cardiac surgery can be detected by measurement of electrical bioimpedance (EBI). Conventional pacing wires were replaced by pacing wires with sputtered iridium coating in order to reduce polarization associated with two-electrode impedance measurements. A custom-built bioimpedance analyzer (Osypka Medical GmbH, Berlin, Germany) measured the real part of impedance Re(Z) and the phase (ϕ) at three frequencies (1, 10, and 1000 kHz) and determined an extracellular space index (EZRI) as the quotient of Re(Z) at 1000 kHz and Re(Z) at 1 kHz. Our study included six patients (conventional coronary artery bypass graft, age 68.1 ± 8.3 years) subject to routine cardioplegic ischemia and reperfusion. Preischemic bioimpedance measurements were not impaired by interference of the beating heart. Intraischemically, bioimpedance at 1 kHz and phase at 10 kHz increased until opening of a bypass graft, which is probably induced by closure of gap junctions and cell swelling processes. After cross clamping, EZRI slowly decreased as an effect of mild cell swelling. After ischemia, values returned almost to baseline measurements, indicating sufficient reperfusion processes. Measurement of EBI correlates with myocardial ischemic injury and is applicable in a two-electrode setup providing low-polarization pacing wires.