Abstract

Composite self-excited PZT-glass cantilevers (5 and 3 mm in length, 1.8 and 2.0 mm wide) were fabricated and their resonance characteristics were determined in air and at 1 mm liquid immersion. In air, resonance occurred at 65.8 and 63.4 kHz for the two cantilevers used in this paper. Monoclonal antibody (MAb) specific to the pathogen Escherichia coli (E. coli) O157:H7 was immobilized at the cantilever glass tip, and then exposed to pathogen in the concentration range of 7x10(2) to 7x10(7)bacteria/mL. Resonance of the second mode decreased due to pathogen attachment in accordance with a proposed kinetic model. The specific attachment rate constant was found to be 3x10(-9) to 5x10(-9) min-1 (cell/mL)-1. Exposure to a mixed population containing both a pathogenic and non-pathogenic strain showed that the antibody-immobilized cantilever is highly selective, thus demonstrating its usefulness for detecting water-borne pathogens.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.