Abstract

Implantable loop recorders (ILRs) are increasingly used in equine cardiology to detect arrhythmias in the context of collapse, poor performance or monitoring for recurrence of atrial fibrillation (AF). However to date, the ILR has never been reported to be used with a remote monitoring functionality in horses, therefore the arrhythmia is only discovered when a clinician interrogates the ILR using dedicated equipment, which might delay diagnosis and intervention. This case report describes the use of an ILR with remote monitoring functionality in a horse with recurrent AF. The remote monitoring consisted of a transmission device located in the stable allowing daily transmission of arrhythmia recordings and functioning messages to an online server, available for the clinician to evaluate without specialised equipment. The ILR detected an episode of paroxysmal AF approximately three months after implantation. Seven months after implantation, initiation of persistent AF was seen on an episode misclassified by the ILR as bradycardia, and the horse was retired. This report shows the feasibility and benefits of remote monitoring for ILRs in horses, but also the shortcomings of current algorithms to interpret the equine electrocardiogram.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call