Sensitive and selective detection of nitric oxide (NO) in the human body is crucial since it has the vital roles in the physiological and pathological processes. This study reports a new type of electrochemical NO biosensor based on zinc-dithiooxamide framework derived porous ZnO nanoparticles and polyterthiophene-rGO composite. By taking advantage of the synergetic effect between ZnO and poly(TTBA-rGO) (TTBA = 3'-(p-benzoic acid)-2,2':5',2″-terthiophene, rGO = reduced graphene oxide) nanocomposite layer, the poly(TTBA-rGO)/ZnO sensor probe displays excellent electrocatalytic activity and explores to detect NO released from normal and cancer cell lines. The ZnO is immobilized on a composite layer of poly(TTBA-rGO). The highly porous ZnO offers a high electrolyte accessible surface area and high ion-electron transport rates that efficiently catalyze the NO reduction reaction. Amperometry with the modified electrode displays highly sensitive response and wide dynamic range of 0.019-76 × 10-6 m with the detection limit of 7.7 ± 0.43 × 10-9 m. The sensor probe is demonstrated to detect NO released from living cells by drug stimulation. The proposed sensor provides a powerful platform for the low detection limit that is feasible for real-time analysis of NO in a biological system.