Abstract
A planar array immunosensor, equipped with a charge-coupled device (CCD) as a detector, was used to simultaneously detect 3 toxic analytes. Wells approximately 2 mm in diameter were formed on glass slides using a photoactivated optical adhesive. Antibodies against staphylococcal enterotoxin B (SEB), ricin, and Yersinia pestis were covalently attached to the bottoms of the circular wells to form the sensing surface. Rectangular wells containing chicken immunoglobulin were used as alignment markers and to generate control signals. After removing the optical adhesive, the slides were mounted over a scientific grade CCD operating at ambient temperature in inverted (multipin phasing) mode. A two-dimensional graded index of refraction lens array was used to focus the sensing surface onto the CCD. Solutions of toxins were then placed on the slide. After rinsing, Cy5-labeled antibodies were introduced. The identity and amount of toxin bound at each location on the slide were determined by quantitative image analysis. Concentrations as low as 25 ng/mL of ricin, 15 ng/mL of pestis F1 antigen, and 5 ng/mL of SEB could be routinely measured.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.