Abstract

The present study was undertaken to determine the usefulness of a new, sensitive urinary dipstick, Micral-Test II (Boehringer-Mannheim, Mannheim, Germany) for the detection of microalbuminuria (urinary albumin concentration: 20–200 mg/1) in diabetic patients. Four hundred and eleven consecutive outpatients with diabetes were screened for microalbuminuria. The diagnostic accuracy of Micral-Test II for the detection of urinary albumin excretion >20 mg/l was assessed by comparing the urinary albumin concentration determined by Micral-Test II with radioimmunoassay (RIA). In comparison with radioimmunoassay, Micral-Test II had an overall sensitivity of 93%, specificity of 93% and positive predictive value of 89%. False negative results were found in 7% and false positives were obtained in 7% of specimens. These findings suggest that Micral-Test II may be a useful method for the sceening of microalbuminuria. However, it should not be regarded as a diagnostic test and a positive result should be followed by measuring AER by a laboratory based method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.