Abstract

LaBr$_3$:Ce crystals, with photomultiplier or single SiPM readout (up to 3x3 mm$^2$ ) have been introduced for radiation imaging in medical physics. An R&D was pursued with 1/2" and 1" LaBr3:Ce crystals, from different producers, to realize compact large area detectors (up to some cm$^2$ area) with SiPM array readout, aiming at high light yields, good energy resolution, good detector linearity and fast time response for low-energy X-rays. The study was triggered by the FAMU experiment at the RIKEN-RAL muon facility, aiming at a precise measurement of the proton Zemach radius to solve the so-called "proton radius puzzle". The goal is the detection of characteristic X-rays around 130 KeV. Other applications may be foreseen homeland security and γ-ray astronomy. Results were obtained with a direct readout based on a CAEN V1730 FADC, with no need for an amplification stage. At the Cs$^{137}$ peak, energy resolutions up to ∼ 3% were obtained, using a readout with Hamamatsu SiPM arrays. These results compare well with best available results obtained with a PMT readout.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.