Abstract

We report one case of loosening of the locking bolt in the stem-condyle junction of a constrained modular femoral component in revision total knee arthroplasty. Early detection of loosening of the locking bolt was possible using reconstructed view of computerized tomography scans before complete disengagement and radiographic loosening of the stemmed femoral component. When using the Total Condyle III femoral component with a stem extension, surgeons should keep in mind that stress concentration at the stem-condyle junction can result in failure of the locking bolt although it is rare and that loosening or disengagement of the locking bolt may be a sign of a loose stemmed femoral component.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.