Abstract
The treatment with high doses of ionizing radiation is prescribed for the terminal sterilization of many healthcare products. In order to check the product irradiation, it may be useful to develop suitable detection methods. The aim of this work was to verify the suitability of the thermoluminescence (TL) and electron paramagnetic resonance (EPR) techniques to reveal illegal omission of radiation sterilization in glass used for blood test tubes. In particular, this work reports the results of the analysis performed on two batches of glass tubes provided by the same manufacturer. The results indicated that both TL and EPR techniques could reveal the damage induced by high doses of ionizing radiation in the glass allowing the identification of the tubes which were irradiated at the manufacture stage. The study provided also some information about the chemical composition of the two-batch glass.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.