Abstract

A comparative evaluation of the following commercial immunoassays for the determination of antibodies to Toxoplasma gondii was performed: Behring Diagnostics OPUS Toxo G and Toxo M, Abbott Diagnostics IMX Toxo-IgG 2.0 and Toxo-IgM, Sanofi Diagnostics Pasteur Platelia Toxo IgG and Toxo IgM, and bioMérieux Vitek VIDAS Toxo IgG and IgM. Of 676 specimens that were tested for Toxoplasma-specific immunoglobulin G (IgG) antibodies, 26% were reactive by all methods while 8% displayed some discrepancy. Of 718 specimens that were tested for Toxoplasma-specific IgM antibodies, 3% were reactive by all methods while 10% displayed some discrepancy. Analysis of discrepant specimens revealed performance shortcomings with all IgM-specific assays. The impact of such shortcomings is magnified in a population with a low prevalence of toxoplasmosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.